Bowie Dick Simulation Test | USA
Process challenge device (PCD) for daily functionality check (type test, no sterility test) of air removal in steam sterilizers after start up, Compact-PCD, light blue
Hospitals & Clinics,
Equipment Manufacturer
Select Indicator Type
Product information
product details
- Functional test (type test, not sterility test) for daily checking of air removal in steam sterilizers after switching on
- The test does not replace ongoing routine monitoring during sterile goods production
- Simulates an air removal test according to AAMI/ANSI79 (4 kg cotton pack), validated according to the test method described in ISO 11140-5
- Indicator strips available as refill pack with 100, 250 and 500 pieces