Bowie Dick Simulation Test

Process challenge device (PCD) for daily functionality check (type test, no sterility test) of air removal and steam penetration in steam sterilizers after start up, Compact-PCD, purple

A small tagging Icon

Hospitals & Clinics,

Dental,

Equipment Manufacturer

Select Product

View Technical Data

Product information

Features

  • Functional test (type test, not sterility test) for daily checking of air removal and steam penetration in steam sterilizers after switching on

  • Mandatory when operating large sterilizers according to EN 285, but does not replace ongoing routine monitoring during sterile goods production
  • Simulates an air removal and steam penetration test according to EN 285 (7 kg cotton pack), validated according to the test method described in EN ISO 11140-4

  • Indicator strips  available as refill pack with 100, 250 and 500 pieces