Bowie Dick Simulation Test | Europe Hollow Load Test
Process challenge device (PCD) for daily functionality check (type test, no sterility test) of air removal and steam penetration in steam sterilizers after start up, Compact-PCD, blue
Hospitals & Clinics,
Dental,
Equipment Manufacturer
Select Indicator Type
Product information
product details
- Functional test (type test, not sterility test) for daily checking of air removal and steam penetration in steam sterilizers after switching on
- Mandatory when operating large sterilizers according to EN 285, but does not replace ongoing routine monitoring during sterile goods production
- Simulates an air removal and steam penetration test according to EN 285 (7 kg cotton pack), validated according to the test method described in EN ISO 11140-4 and at the same time meets the requirements of a helix hollow load test according to EN ISO 11140-6 (combination of BD and hollow load test in one test system)
- Indicator strips available as refill pack with 100, 250 and 500 pieces